Supplement to the instructions for use Oxylog 3000 plus 3
English
Problem solving
Messages in the alarm message field
Technical data
Device specifications
Alarm Cause Remedy Rank
! CO
2 inac-
curate or
failed
The CO
2 measure-
ment is inaccurate.
Use external CO2
monitoring.
Contact
DrägerService.
42
The CO
2 sensor is
defective.
Replace the faulty
CO
2 sensor.
The zero calibra-
tion or filter check
of the CO
2 sensor
was activated and
no valid measured
values were gener-
ated.
Wait until the CO
2
zero calibration/fil-
ter check is com-
pleted.
!! etCO
2
low
The lower alarm
limit for the end-
tidal CO
2 concen-
tration has been
exceeded.
Check patient con-
dition.
27
Check alarm limits.
Adjust the alarm
limit, if necessary.
The CO
2 sensor
has fallen off the
cuvette.
Reattach the CO2
sensor to the
cuvette.
! No int.
battery
charging
Internal battery
cannot be charged
due to a faulty bat-
tery or too hot or
cold environment.
Press the Alarm
Reset key to con-
firm the alarm.
Remove the inter-
nal battery and
reinsert or replace
it.
44
Protection class
against ingress of
liquids
IP34, the device is splash-proof from all
directions and protected against foreign
objects >2.5 mm (0.1 in)