112 Instructions for use | Oxylog VE300 SW 1.n
Technical data
Degree of protection of the CO2 sensor
Defibrillation recovery time
15.3 Ambient conditions
During operation
During storage and transportation
Ventilator without battery, with reusable breathing circuit
Disposable breathing circuit
IP64
0 seconds
Temperature (device) –20 to +50 °C (–4 to +122 °F)
Temperature (power supply unit) 0 to +50 °C (+32 to +122 °F)
Permissible operating tempera-
ture during charging
0 to +45 °C (+32 to +113 °F)
Ambient pressure (device) 620 to 1100 hPa (8.99 to 15.95 psi)
Automatic ambient pressure compensation
within this pressure range
Height above sea level up to 4000 m (13123 ft)
Ambient pressure (power supply
unit)
700 to 1060 hPa (10.15 to 15.37 psi)
Relative humidity 5 to 95 % (without condensation)
Temperature –40 to +70 °C (–40 to +158 °F)
Ambient pressure 620 to 1100 hPa (8.99 to 15.95 psi)
Relative humidity 5 to 95 % (without condensation)
Time required for the device tem-
perature to rise from the mini-
mum storage temperature to the
minimum operating temperature
40 minutes
Time required for the device tem-
perature to rise from the maxi-
mum storage temperature to the
maximum operating temperature
66 minutes
Temperature –20 to +50 °C (–4 to +122 °F)
Ambient pressure 570 to 1200 hPa (8.27 to 17.40 psi)
Relative humidity 5 to 95 % (without condensation)