32 Instructions for use | Oxylog VE300 SW 1.n
Operating concept
4.3 Color concept
Values can be selected by touching the screen, and can be changed and confirmed
using the rotary knob.
Colors indicate the availability of functions and settings on therapy controls and
buttons.
4.4 Selecting and setting
The settings always require confirmation by pressing the rotary knob.
1. Select
Touch the control element. The color changes to yellow.
2. Set
Turn the rotary knob to adjust a value.
3. Confirm
Press the rotary knob to confirm the value. The color changes to dark green.
When the value has been selected and changed, the message To confirm the
setting, press the rotary knob. is also displayed in the notification field. This
message starts to flash after a short time.
Canceling the setting or changing procedure
If the parameter shall not be changed (color still yellow), the following options allow
the existing settings to be kept:
● Touch the changed parameter again. This resets the parameter selection and
change.
● Select a different parameter. This resets the previously selected parameter to its
original value.
● Do not press the rotary knob. The value of the parameter resets to the previous
setting after a few seconds.
● Touch another area of the touchscreen.
Color Example Meaning
Dark green Operable element: function activated
Yellow Selected element: not yet confirmed with
rotary knob