Instructions for use | Oxylog VE300 SW 1.n 37
Assembly and preparation
The device cannot use the input voltage outside the range. The on-board power
supply must be protected by a suitable 10 to 16 A DC fuse.
Mount the DC/DC converter on a flat wall and make sure that the wall is solid
enough to support the bracket. Use all 4 mounting holes (screw size M4).
1. Plug the large connector of the DC/DC converter into the on-board power
supply.
2. Plug the small connector into the device's DC socket.
3. Check if the device is connected correctly to the external power source.
– The LED next to the battery status indicator lights up green when the battery
is charged.
– The LED next to the battery status indicator flashes green when the battery is
charging.
– The LED next to the battery status indicator lights up yellow if the battery is
faulty, or no battery is inserted.
External power supply with a power supply unit
WARNING
Risk of patient injury and damage to the device
When using the power supply unit outdoors, there is a risk of electric shock and
damage to the device.
► Do not use the power supply unit outdoors.
► Do not touch the power supply unit and the patient at the same time.
1. Insert the power plug into the power socket.
2. Connect the DC connector to the device's DC socket.
3. Check if the device is connected correctly to the external power source.
– The LED next to the battery status indicator lights up green when the battery
is charged.
– The LED next to the battery status indicator flashes green when the battery is
charging.
– The LED next to the battery status indicator lights up yellow if the battery is
faulty, or no battery is inserted.
To disconnect the ventilator from the power supply, disconnect the power cable
from the power socket.
5.6 Connecting the gas supply
Special care must be taken when handling oxygen.
► For further information, see the following chapter: "Use of oxygen", page 13.
The device can be operated with 100 % and 93 % medical oxygen.
The oxygen concentration in the breathing gas will be reduced when using 93 %
oxygen. For further information, see the following chapter: "Device
settings", page 81.
The device supports this by allowing the oxygen content display to be configured
accordingly.