38 Instructions for use | Oxylog VE300 SW 1.n
Assembly and preparation
Connecting an external oxygen cylinder
WARNING
Risk of ventilator malfunction
Components installed in the gas supply line may cause device malfunctions.
Do not install dosage valves or flowmeters in the device's gas supply line.
To avoid insufficient oxygen supply during use, always check the oxygen
cylinder pressure before use.
1. Connect the pressure reducer (B) to the oxygen cylinder (A).
2. Connect the oxygen compressed gas hose (C) to the standard port (D) on the
device.
3. Connect the oxygen compressed gas hose (C) to the pressure reducer (B).
4. Rotate the cylinder valve slowly and open it fully.