42 Instructions for use | Oxylog VE300 SW 1.n
Assembly and preparation
1. Place the diaphragm (B) in the breathing valve housing (C). Make sure that it is
fitted correctly.
2. Mount the cover (A) and turn it about 60° clockwise until it engages.
3. Push the pressure sensor (D) onto the breathing valve (C). To do so, align the
pressure sensor (D) on the breathing valve using the grooves in the pressure
sensor.
4. Push the elbow onto the pressure sensor (D).
Connecting the breathing hose and the pressure measuring line
1. Connect the breathing hose (A) to the breathing valve.
2. Connect the pressure measuring line (B) to the connector of the pressure
sensor.
3. Connect the pressure measuring line to the device.
4. Connect the breathing hose to the device's gas outlet.
3025230253
A
B
C
D
E
A
B