46 Instructions for use | Oxylog VE300 SW 1.n
Assembly and preparation
6. Insert the CO2 sensor cable in the cable clips on the breathing hose.
Alternatively, connect the CO
2 cuvette directly to the patient port of the elbow,
without disconnecting the elbow from the flow sensor.
After connecting the CO
2 sensor (mainstream measurement, PGM algorithm,
atmospheric pressure compensation) or after a power supply failure, the sensor
must warm up for approx. 3 minutes before it is ready for operation. Pay attention to
the following during the warm-up phase and at temperatures below 10 °C (50 °F):
–The etCO
2 and CO2 values may have a reduced accuracy.
The CO
2 sensor cable may be extended with a maximum of one extension cable.
For further information, see the following chapter: "List of accessories", page 125.
5.10 Attaching the device to rail systems
The device can be attached to various rail systems or to stretcher bars with
diameters up to 35 mm (1.4 in) by means of a rail holder (optional).
The distance from the stretcher bar or rail to the wall or to a supporting stretcher
bar must not exceed 30 mm (1.2 in).
Attach the device in a suitable position orientated to the patient. The device is
held by its own weight.
WARNING
Risk of patient injury and damage to the device
The device might come loose and drop when being transported with the rail holder.
Observe wall clearance requirements.
Secure or protect the device by additional means when transporting it.