Instructions for use | Oxylog VE300 SW 1.n 49
Getting started
6.3 Performing the system tests
The device has 2 different system tests:
System test
System test with breathing circuit
Both system tests can only be performed in standby mode. In operation mode the
buttons to start the system tests are grayed-out, and a relevant advisory message
is displayed.
6.3.1 System test
A system test must be performed in the following cases:
Once a day, e.g., at the start of a shift
At least every 6 months
The system test comprises the following single tests:
Pneumatic test
Leakage test
The system test takes about 1 minute.
6.3.1.1 Performing the system test
1. To switch the ventilator on, briefly press the start/standby key.
The ventilator performs an automatic self-test, then the standby screen is
displayed.
2. To move to screen 2 of the dialog, touch the button 1 in the top right-hand corner
of the screen.
3. Touch the button Test & calibration (A).
4. Touch the button System test (B).
2937530183
A
B