50 Instructions for use | Oxylog VE300 SW 1.n
Getting started
The system test can be canceled at any time by touching the X button in the
top right-hand corner of the screen.
5. On screen 1/4 confirm the advisory messages 1. Make sure that no patient is
connected. and 2. Connect the O
2 supply. with the Next button.
6. On screen 2/4 confirm the advisory message 3. Make sure that the breathing
circuit is disconnected. with the Next button.
The system test is then started. Its progress is indicated by a bar on screen 3/4.
7. When prompted Seal the gas outlet., seal the device port to which the
breathing circuit is to be connected.
The test result is displayed on screen 4/4.
8. To exit, touch the Finish button in the bottom right-hand corner of the screen.
6.3.2 System test with breathing circuit
A system test with breathing circuit must be performed in the following cases:
After every reprocessing of the breathing circuit
At least every 6 months
The system test with breathing circuit comprises the following single tests:
Pneumatic test
Breathing circuit test
Leakage test
The system test with breathing circuit takes about 1 minute.
6.3.2.1 Performing the system test with breathing circuit
1. To switch the ventilator on, briefly press the start/standby key.
The ventilator performs an automatic self-test, then the standby screen is
displayed.
2. To move to screen 2 of the dialog, touch the button 1 in the top right-hand corner
of the screen.
3. Touch the button Test & calibration (A).
29375
A