EasyManua.ls Logo

Dräger Oxylog VE300 - Page 53

Dräger Oxylog VE300
134 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use | Oxylog VE300 SW 1.n 53
Getting started
7. On screen 1/3 confirm the advisory message Remove the cuvette from the
CO
2 sensor. with the Next button.
The CO
2 zero calibration starts and the message Zero calibration of CO2
sensor in progress... is displayed. A bar indicates that the calibration is in
progress.
When the CO
2 zero calibration is completed, a message is displayed indicating
the result of the calibration.
8. To exit, touch the Finish button in the bottom right-hand corner of the screen.
9. Reconnect the CO
2 sensor to the CO2 cuvette of the breathing circuit.
If CO
2 zero calibration was not successful:
The screen displays the alarm message CO
2 zero calibration failed.
Repeat the CO
2 zero calibration.
If CO
2 zero calibration is still not successful:
1. Check whether the CO
2 sensor is soiled, and clean the CO2 sensor if necessary.
If the sensor is faulty, replace the sensor.
2. Repeat the CO
2 zero calibration.
6.4.2 Performing the CO2 filter test (option)
1. To switch the ventilator on, briefly press the start/standby key.
The ventilator performs an automatic self-test, then the standby screen is
displayed.
2. Connect the CO
2 sensor to the device.
3. To move to screen 2 of the dialog, touch the button 1 in the top right-hand corner
of the screen.
4. Touch the button Test & calibration (A).
5. Touch the button CO
2 filter test (B).
2937533176
A

Table of Contents

Other manuals for Dräger Oxylog VE300

Related product manuals