54 Instructions for use | Oxylog VE300 SW 1.n
Getting started
The CO2 filter test can be canceled at any time by touching the X button in the
top right-hand corner of the screen.
Shortly after connecting the CO2 sensor, the filter test may take longer (up to
2 minutes).
6. Disconnect the CO
2 sensor (A) from the CO2 cuvette and connect it to the test
filter (B).
7. On screen 1/3 confirm the advisory messages 1. Remove the cuvette from the
CO
2 sensor. and 2. Attach the CO2 sensor to the test filter. with the Next
button.
8. The CO
2 filter test starts and the message Filter test of CO2 sensor in
progress... is displayed. A bar indicates that the CO
2 filter test is in progress.
9. When the CO
2 filter test is completed, a message is displayed indicating the
result.
10.To exit, touch the Finish button in the bottom right-hand corner of the screen.
11. Reconnect the CO
2 sensor to the CO2 cuvette.
If the CO
2 filter test was not successful:
The screen displays the alarm message CO
2 filter test failed. The measured CO2
value is outside the permissible tolerance range.
1. Check whether the CO
2 sensor or the test filter is soiled, and clean it if
necessary.
2. Repeat the CO
2 filter test.
30255