Instructions for use | Oxylog VE300 SW 1.n 57
Operation
1. Adjust the following parameters using the therapy controls:
A Respiratory rate RR
B Tidal volume VT
2. Press the More settings button (C) and adjust the following parameters and
settings:
A Inspiratory oxygen concentration FiO
2 (100 % O2 or O2/air mix)
B Maximum airway pressure Pmax
C Positive end-expiratory pressure PEEP
D Ratio of inspiratory time to expiratory time I:E
E Trigger sensitivity Trigger
When setting the ratio of inspiratory time to expiratory time I:E the corresponding
value Ti is automatically displayed in the notification field.
VC-AC
Volume-Controlled - Assist Control
For synchronization with the patient's inspiratory efforts. For patients with partial
spontaneous breathing.
If a trigger value is set in ventilation mode VC-CMV, the ventilation mode
automatically changes to VC-AC.
When the trigger sensitivity is activated and set, the mandatory breaths are
synchronized with the patient's inspiratory efforts.
A low value corresponds to a high sensitivity. To adapt the trigger sensitivity to the
patient, increase the value if triggering is too easy (Autotrigger) and reduce it if
triggering is too difficult.
The actual respiratory rate may be higher than the set respiratory rate RR.
Successful patient triggering is briefly indicated by an asterisk (*) in the middle of
the waveform field at the top.