Instructions for use | Oxylog VE300 SW 1.n 67
Operation
O2 concentration which can be realized at a mean airway pressure of 60 mbar
(60 cmH
2O)
When using 93 % oxygen as the supply gas, the actual O2 concentration is
slightly lower than the indicated values.
WARNING
Risk of patient injury
Strictly observe the following information regarding toxic and infectious
environments:
► The patient must be ventilated with an inspiratory oxygen concentration of
FiO
2 100 % O2 or 93 % O2 so that toxic and infectious constituents cannot enter
the breathing gas.
► The patient must immediately be transferred to a breathable atmosphere in
order to prevent inhalation of toxic and infectious air when spontaneous
breathing resumes.
► Use a bacteria filter.
7.7 Switching to standby mode and turning off the device
After disconnecting the patient, switch the ventilator to standby mode or turn it off.
To switch to standby mode:
1. Press the start/standby key (A) for about 3 seconds.
2. Touch the Standby button and confirm.
The standby screen is then displayed. For further information, see the following
chapter: "Structure of the screen", page 29.