82 Instructions for use | Oxylog VE300 SW 1.n
Configuration
– Buttons Back, Next and Finish (G) for navigating within the dialog.
On dialog page 2/2 the following system settings can be displayed and changed:
– Alarm volume (H)
– 25 %, 50 %, 75 %, 100 %
– Min. alarm volume (I)
– 25 %, 50 %, 75 %, 100 %
– Screen brightness (J)
– 25 %, 50 %, 75 %, 100 %
– Screenshots (K)
– On, Off
– Bluetooth (L)
– On, Off
– Buttons Back, Next and Finish (D) for navigating within the dialog.
9.9.8 Ventilation settings
Here the ventilation settings with which the device starts up can be specified.
WARNING
Risk of patient injury
When using several ventilators in the same environment of use, the ventilation
settings must be the same on all ventilators to prevent risk of patient injury.
► Make sure that all ventilators have the same ventilation settings and alarm
settings.
Start mode can be selected on the first two dialog pages of the ventilation settings.
33062
33065