92 Instructions for use | Oxylog VE300 SW 1.n
Troubleshooting
! Device
malfunction
Technical fault that is
not related to the
patient settings.
To continue ventilation
with this device, contin-
uously monitor the
patient.
Perform the system
test.
Contact DrägerService.
45
!!! Device tem-
perature high
The device temperature
is above the maximum
operating temperature.
Move the device to a
cooler environment.
13
!!! Device tem-
perature low
The device temperature
is below the minimum
operating temperature.
Move the device to a
warmer environment.
14
!!! Device tem-
perature very
high
The device temperature
is well above the maxi-
mum operating tem-
perature.
Urgently move the
device to a cooler envi-
ronment.
3
!! etCO
2 high The end-tidal CO2 con-
centration is above the
upper alarm limit.
Check the patient's
condition.
34
Check the alarm limits.
Adjust the alarm limit if
necessary.
!! etCO
2 low The end-tidal CO2 con-
centration is below the
lower alarm limit.
Check the patient's
condition.
35
Check the alarm limits.
Adjust the alarm limits if
necessary.
!!! Expiratory
minute
volume high
The minute volume
MVe is above the upper
alarm limit.
Check the patient's
condition, check the
ventilation settings,
adjust the alarm limits if
necessary.
17
The flow sensor is
faulty.
Replace the flow sen-
sor.
Technical fault. Disconnect the device
from the patient and
continue ventilation
immediately using
another ventilator.
Contact DrägerService.
Alarm
prior-
ity
Alarm Cause Remedy Inter-
nal pri-
ority