Instructions for use | Oxylog VE300 SW 1.n 93
Troubleshooting
!!! Expiratory
minute vol-
ume low
The minute volume
MVe is below the lower
alarm limit.
Check the patient's
condition, check the
ventilation settings,
adjust the alarm limits if
necessary.
16
There is leakage in the
breathing circuit.
Make sure that the con-
nections in the breath-
ing circuit are tight.
The flow sensor is
faulty.
Replace the flow sen-
sor.
Technical fault. Disconnect the device
from the patient and
continue ventilation
immediately using
another ventilator.
Contact DrägerService.
!!! Expiratory
pressure high
The breathing valve or
the breathing circuit is
obstructed.
Check the patient's
condition.
Check the breathing
valve and the breath-
ing circuit.
10
The expiratory resis-
tance is increased.
Check the bacteria filter
or the HME filter and
replace it if necessary.
Technical fault. Disconnect the device
from the patient and
continue ventilation
immediately using
another ventilator.
Contact DrägerService.
!!! Forced
shutdown
The start/standby key
has been continuously
pressed for a few sec-
onds.
Release the
start/standby key or
press it briefly.
1
Technical fault. Disconnect the device
from the patient and
continue ventilation
immediately using
another ventilator.
Contact DrägerService.
Alarm
prior-
ity
Alarm Cause Remedy Inter-
nal pri-
ority