96 Instructions for use | Oxylog VE300 SW 1.n
Troubleshooting
11.3 Messages in the notification field
Message Cause Explanation/Remedy
Bluetooth PIN: The Bluetooth function was
activated.
Display of the PIN that
must be used as the code
by an external Bluetooth
device.
I:E or Psupp or Ti A ventilation parameter is
adjusted.
The additional parameter is
displayed as help. Set the
desired value for the venti-
lation parameter and con-
firm with the rotary knob.
Maximum flow
reached, check I:E
The flow resulting from the
setting of the tidal volume
VT relative to Ti or I:E is not
achievable.
Change tidal volume VT or
inspiratoy time Ti or breath-
ing time ratio I:E or respira-
tory rate RR.
PEEP > 10? Confirm
with rotary knob.
PEEP is set to a value
below 10 mbar
(10 cmH
2O).
To confirm the message,
press the rotary knob, then
set the desired target value
and confirm again with the
rotary knob.
Pinsp ≥ PEEP + 3 Pinsp or PEEP is changed. Set Pinsp >
PEEP +3 mbar.
Pmax ≥ PEEP +
∆Psupp
When adjusting PEEP or
∆Psupp, the upper limit
Pmax is reached.
Select the settings so that
the sum of PEEP and
∆Psupp is below Pmax.
Pmax ≥ Pinsp Pmax or Pinsp is changed. Set Pmax > Pinsp.
RR too high, check Ti The inspiratory and/or expi-
ratory times resulting from
the settings for RR and I:E
or Ti are not achievable.
Change Ti, I:E, or RR.
Screenshot saved The screenshot symbol was
touched.
The screenshot was saved
in the internal device mem-
ory.
Ti too high, check RR The inspiratory and/or expi-
ratory times resulting from
the settings for RR and I:E
or Ti are not achievable.
Change Ti, I:E, or RR.
To confirm the alarm
setting, press the
rotary knob.
An alarm limit is adjusted. Press the rotary knob to
confirm.
To confirm the selec-
tion, press the rotary
knob.
Ventilation is stopped. Press the rotary knob to
confirm.
To confirm the setting,
press the rotary knob.
A ventilation parameter is
adjusted.
Press the rotary knob to
confirm.