96 Instructions for use | Oxylog VE300 SW 1.n
Troubleshooting
11.3 Messages in the notification field
Message Cause Explanation/Remedy
Bluetooth PIN: The Bluetooth function was
activated.
Display of the PIN that
must be used as the code
by an external Bluetooth
device.
I:E or Psupp or Ti A ventilation parameter is
adjusted.
The additional parameter is
displayed as help. Set the
desired value for the venti-
lation parameter and con-
firm with the rotary knob.
Maximum flow
reached, check I:E
The flow resulting from the
setting of the tidal volume
VT relative to Ti or I:E is not
achievable.
Change tidal volume VT or
inspiratoy time Ti or breath-
ing time ratio I:E or respira-
tory rate RR.
PEEP > 10? Confirm
with rotary knob.
PEEP is set to a value
below 10 mbar
(10 cmH
2O).
To confirm the message,
press the rotary knob, then
set the desired target value
and confirm again with the
rotary knob.
Pinsp PEEP + 3 Pinsp or PEEP is changed. Set Pinsp >
PEEP +3 mbar.
Pmax PEEP +
Psupp
When adjusting PEEP or
Psupp, the upper limit
Pmax is reached.
Select the settings so that
the sum of PEEP and
Psupp is below Pmax.
Pmax Pinsp Pmax or Pinsp is changed. Set Pmax > Pinsp.
RR too high, check Ti The inspiratory and/or expi-
ratory times resulting from
the settings for RR and I:E
or Ti are not achievable.
Change Ti, I:E, or RR.
Screenshot saved The screenshot symbol was
touched.
The screenshot was saved
in the internal device mem-
ory.
Ti too high, check RR The inspiratory and/or expi-
ratory times resulting from
the settings for RR and I:E
or Ti are not achievable.
Change Ti, I:E, or RR.
To confirm the alarm
setting, press the
rotary knob.
An alarm limit is adjusted. Press the rotary knob to
confirm.
To confirm the selec-
tion, press the rotary
knob.
Ventilation is stopped. Press the rotary knob to
confirm.
To confirm the setting,
press the rotary knob.
A ventilation parameter is
adjusted.
Press the rotary knob to
confirm.