Reprocessing
3. Disconnect the pressure sensor (A) from the breathing valve (B).
4. Carefully disconnect the pressure measuring line (C) from the pressure
sensor (A).
5. Disconnect the elbow (D) from the pressure sensor (A).
6. Disconnect the breathing hose (E) from the breathing valve (B).
Disassembling the reusable breathing circuit for adults (Plus option)
1. Disconnect the breathing hose from the device.
2. Disconnect the flow measuring line or pressure measuring line from the device
ports.
3. Disconnect the flow sensor (A) from the breathing valve (B).
4. Carefully and evenly disconnect the flow measuring line or the pressure
measuring line (C) from the flow sensor (A).
CAUTION
Damage to the device
The flow sensor may be damaged when disconnecting the flow measuring line
or the pressure measuring line from the flow sensor ports.
► Never twist the flow measuring line or the pressure measuring line and do not
exert force on them.
3325030171