20 Reprocessing instructions | Perseus A500
Reprocessing
3.5 Validated reprocessing procedures
3.5.1 Overview of the reprocessing procedures of the components
Components Surface disinfec-
tion with cleaning
Machine cleaning
with thermal dis-
infection
Description of the
procedure
Device surfaces
Yes No
(see "Surface disin-
fection with clean-
ing", page 21)
Breathing system
mount
(see "Surface disin-
fection with clean-
ing", page 21)
Upper part of the
breathing system
and lower part of
the breathing sys-
tem
No Yes
(see "Upper part of
the breathing sys-
tem", page 25)
Inspiratory port and
expiratory port
(see "Inspiratory
port and expiratory
port", page 26)
Inspiratory valve
(red) and expiratory
valve (blue)
(see "Inspiratory
valve and expira-
tory valve",
page 27)
APL valve (see "APL valve",
page 26)
Absorber container
and absorber insert
(see "Absorber
insert", page 28)
TurboVent 2 blower
module
(see "TurboVent 2
blower module",
page 29)
Flexible breathing
bag arm (option)
Yes
1)
1) Cleaning with thermal disinfection is required for effective, hygienic reprocessing
Yes
(see "Flexible
breathing bag arm