Loading...
Checking the lighting
1 Press the light switch
M
:
The resting surface of the bed will be illuminated.
1 Press switch
M
once again:
The light will be switched off.
2 Press the power switch in as far as it will go = OFF.
Checking the skin temperature sensor
Only for skin temperature control.
With the apparatus switched on,
3 Press the
µ
key: the apparatus switches over to
»Man« (manual) mode.
4 Plug in the skin temperature sensor.
5 The measured temperature is displayed,
or
3 dashes at bottom of display = temperature
less than 33 °C
3 dashes at top of display = temperature
above 38 °C
If 3 dashes are displayed in the middle of the display:
See section on "Troubleshooting", page 50.
Checking that the care devices are ready for
operation
Check the Babytherm 8000 WB and Photo-
Therapy 4000 as indicated in their specific
Instructions for Use.
Preparation
Checking readiness for operation
36