EasyManua.ls Logo

Dräger Vapor 2000 - Test anaesthetic for extraneous liquids; Test anaesthetic properties; Test handwheel; Test keyed filling system

Dräger Vapor 2000
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
This test list does not replace the inspections and servicing to be carried out by DrägerMedical AG & Co. KGaA
5327.400 Vapor 2000 04.99 Test list Page 23
GBK5327400TL1.fm 19.01.00
Copyright reserved. Reproduction permitted only for non-commercial purposes.
2 General
3Anaesthetic
3.1 Test anaesthetic for extraneous liquids
Make sure the vaporizer chamber contains only the intended anaesthetic.
Make sure the anaesthetic is not intermixed with water or any other liquid.
3.2 Test anaesthetic properties
Make sure the anaesthetic is not contaminated (by particles or suspen-
sions) or discoloured.
If necessary, flush out the vaporizer with the specified anaesthetic.
Test the relevant anaesthetic vaporizer, type
Dräger-Vapor 2000 (from now on referred to as the “Vapor”)
before testing the associated anaesthetic machine.
If you use a Dräger Irina or Iria unit for anaesthetic
concentration measurement, allow for the warm-up phase
of the sensors. Switch on the relevant unit before making
the check.
If there is water or any other liquid in the vaporizer chamber,
the vaporizer must be subjected to a general overhaul at the
manufacturers.

Related product manuals