This test list does not replace the inspections and servicing to be carried out by DrägerMedical AG & Co. KGaA
5327.400 Vapor 2000 04.99 Test list Page 24
GBK5327400TL1.fm 19.01.00
Copyright reserved. Reproduction permitted only for non-commercial purposes.
4 Housing, labels and functions
4.1 Test housing and labels
• Check that the following vaporizer components are complete and
undamaged:
− The plug-in adapter
− The vaporizer housing
− The handwheel with adjuster cap
− The sightglass with its cover
− The front caps
• Check that the labelling of the vaporizer is complete and legible.
4.2 Test distinguishing features of anaesthetic
• Check that the handwheel cap is seated firmly on the handwheel.
• If an interlock system 2 is fitted on the vaporizer:
Check that the lugs in both openings are present and undamaged.
• Check that the anaesthetic-specific labelling on the upper front cap
corresponds to the typeplate:
• Check that the colour code of the handwheel cap and the handwheel
match the anaesthetic designation on the typeplate:
Vaporizer type
Halothane
Enflurane
Isoflurane
Sevoflurane
Vaporizer type Colour code
Halothane red
Enflurane orange
Isoflurane purple
Sevoflurane yellow