This test list does not replace the inspections and servicing to be carried out by DrägerMedical AG & Co. KGaA
5327.400 Vapor 2000 04.99 Test list Page 24
GBK5327400TL1.fm 19.01.00
Copyright reserved. Reproduction permitted only for non-commercial purposes.
4 Housing, labels and functions
4.1 Test housing and labels
Check that the following vaporizer components are complete and
undamaged:
The plug-in adapter
The vaporizer housing
The handwheel with adjuster cap
The sightglass with its cover
The front caps
Check that the labelling of the vaporizer is complete and legible.
4.2 Test distinguishing features of anaesthetic
Check that the handwheel cap is seated firmly on the handwheel.
If an interlock system 2 is fitted on the vaporizer:
Check that the lugs in both openings are present and undamaged.
Check that the anaesthetic-specific labelling on the upper front cap
corresponds to the typeplate:
Check that the colour code of the handwheel cap and the handwheel
match the anaesthetic designation on the typeplate:
Vaporizer type
Halothane
Enflurane
Isoflurane
Sevoflurane
Vaporizer type Colour code
Halothane red
Enflurane orange
Isoflurane purple
Sevoflurane yellow