This test list does not replace the inspections and servicing to be carried out by DrägerMedical AG & Co. KGaA
5327.400 Vapor 2000 04.99 Test list Page 25
GBK5327400TL1.fm 19.01.00
Copyright reserved. Reproduction permitted only for non-commercial purposes.
4.3 Test handwheel
Check that the transport locking element and the zero locking element
are functioning precisely.
Check that the handwheel is easily moved over the following ranges:
4.4 Test keyed filling system
Check that the button on the lock catch and the keyed filler adapter are
colour-coded in accordance with the anaesthetic:
4.5 Test “Quick Fil” filling system
Check that the O-ring or the sealing ring is present in the cap and
undamaged.
4.6 Test funnel filling system
Check that the O-ring or the sealing ring is present in the cap and
undamaged.
Ranges: T...0
0...0.2
0.2...full-scale value
Vaporizer type Colour code
Halothane red
Enflurane orange
Isoflurane purple
Sevoflurane yellow
The funnel filling system is not permitted in the European
Economic Area and in EU member states. Vaporizers with
funnel filling systems have no CE mark!