This test list does not replace the inspections and servicing to be carried out by DrägerMedical AG & Co. KGaA
5327.400 Vapor 2000 04.99 Test list Page 27
GBK5327400TL1.fm 19.01.00
Copyright reserved. Reproduction permitted only for non-commercial purposes.
5 Plug-in adapter
Either the DW-2000 plug-in adapter (white version) or the S-2000 plug-in
adapter (Selectatec, white version) may be fitted. Select the relevant test for
your plug-in adapter.
5.1 Plug-in adapter DW-2000 (white version)
5.1.1 Test code ID
• Check that the code ID of the fitted plug-in adapter matches the anaes-
thetic used (see following four diagrams):
No Dräger plug-in adapter from Vapor 19.n (colour: sil-
ver) or Selectatec plug-in adapter (colour: grey) may be
fitted to the Vapor 2000 !
No Dräger 2000 plug-in adapter with milled corners or
S 2000 plug-in adapter may be fitted to the Vapor 2000
with handwheel cap NMD (control dial cap 1)!
Plug-in adapter DW 2000 Code ID:
h for halothane
e for enflurane
i for isoflurane