EasyManua.ls Logo

Dräger Vapor 2000 - Test interlock system 1 for built-on vaporizers; Test interlock system 2 for plug-in vaporizers

Dräger Vapor 2000
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
This test list does not replace the inspections and servicing to be carried out by DrägerMedical AG & Co. KGaA
5327.400 Vapor 2000 04.99 Test list Page 27
GBK5327400TL1.fm 19.01.00
Copyright reserved. Reproduction permitted only for non-commercial purposes.
5 Plug-in adapter
Either the DW-2000 plug-in adapter (white version) or the S-2000 plug-in
adapter (Selectatec, white version) may be fitted. Select the relevant test for
your plug-in adapter.
5.1 Plug-in adapter DW-2000 (white version)
5.1.1 Test code ID
Check that the code ID of the fitted plug-in adapter matches the anaes-
thetic used (see following four diagrams):
No Dräger plug-in adapter from Vapor 19.n (colour: sil-
ver) or Selectatec plug-in adapter (colour: grey) may be
fitted to the Vapor 2000 !
No Dräger 2000 plug-in adapter with milled corners or
S 2000 plug-in adapter may be fitted to the Vapor 2000
with handwheel cap NMD (control dial cap 1)!
Plug-in adapter DW 2000 Code ID:
h for halothane
e for enflurane
i for isoflurane

Related product manuals