This test list does not replace the inspections and servicing to be carried out by DrägerMedical AG & Co. KGaA
5327.400 Vapor 2000 04.99 Test list Page 29
GBK5327400TL1.fm 19.01.00
Copyright reserved. Reproduction permitted only for non-commercial purposes.
6Interlock system
Either interlock system 1, interlock system 2 or the ISO connector can be
fitted to the vaporizer. Select the relevant test for your component.
6.1 Test interlock system 1 for built-on vaporizers
• Check that the interlock works properly and that two-way operation with
the vaporizers is safeguarded. Can you only activate one vaporizer at a
time?
6.2 Test interlock system 2 for plug-in vaporizers
• Check that the interlock works properly and that two-way operation with
plugged-in vaporizers is safeguarded. Can you only activate one vaporizer
at a time?
• Exchange the vaporizers and repeat the test.
• Check that you can only activate one vaporizer at a time.