5327.400 Vapor 2000 05/99 Fault-Cause-Remedy Page 34
For internal use only. Copyright reserved.
GBK5327400F01.fm 19.01.00
Fault Cause Remedy
1 Fault-Cause-Remedy
Fault Cause Remedy
Operation:
No delivered concentration
or delivered concentration
too high/too low.
Vaporizer not filled, vaporizer
empty.
Fill vaporizer, check
concentration.
Handwheel set to “0” or “T” Set handwheel to >
0.2
vol.%.
No vaporizer connected or
one of several ports
unoccupied and open
Connect vaporizer or close
off open port with vaporizer
or direct gas connection
Vaporizer tilted during or
before operation while
handwheel was not set to
“T”.
This may cause liquid
anaesthetic to enter the flow
control valve.
Prior to operation: Flush and
check concentration.
Vaporizer filled with wrong
anaesthetic or anaesthetic
mixture.
Drain and blow out
vaporizer.
Gas flowing through
vaporizer in wrong direction.
Check connecting system.
Leakage, e.g. because plug-
in adapter not tight on seals.
Remove vaporizer, check
plug-in connector interlock
and sealing rings, refit.
Leak test with vaporizer in
handwheel position “0” or
>
0.2 vol.%.
Leakage at port, e.g.
because vaporizer
connected to SA2 or
Titus not converted for
Vapor 2000 use.
Conversion by
DrägerService.