Operation
Calling the Calibration Menu
The calibration menu can only be
accessed by entering a password.
Password on delivery: » 001 «
The default password on delivery can be
changed using the PC software "CC
Vision".
In measuring mode, press the key for
at least 5 seconds.
The function for entering the password is
selected.
The special symbol » « (for the "enter
password" function) is displayed.
The display shows » 000 « with the first
digit flashing.
Use the key to set the flashing digit.
Press the key, the second digit starts
flashing.
Use the key to set the flashing digit.
Press the key, the third digit starts
flashing.
Use the key to set the flashing digit.
Press the key to confirm the password
once it has been set completely.
The calibration menu functions can now
be selected by pressing the key.
Press the key to call the selected
function.