19
Identifying Alarms
Battery pre-alarm
Acknowledging the pre-alarm:
Press the key. Only the audible alarm and the vibration alarm are switched off.
The battery still lasts approx. 20 minutes after the first battery pre-alarm.
Battery main alarm
The battery main alarm cannot be acknowledged or cancelled:
The device is automatically switched off again after 10 seconds.
Device alarm
The device or one or several sensor channels are not ready for operation.
For remedies, refer to page 22 to page 24.
If necessary, commission the Dräger Safety Service Center to eliminate the error.
End of operation
Only in the case of X-am 1100 and X-am 1700.
A warning period starts before the end of the operating time.
After switching on the device, the special symbol » « (for warning) is displayed.
Acknowledging the end of operation warning:
Press the key. The device will continue to operate until the operating time has elapsed.
When the operating time has elapsed, the text » d 0 « flashes. The device cannot be
switched on anymore.
The alarm is indicated by an intermittent alarm message:
Flashing special symbol » « on the right side of the display:
The alarm is indicated by an intermittent alarm message:
Flashing special symbol » « on the right side of the display:
Intermittent alarm message:
Special symbol » « on the right side of the display:
OK
on
OK
on

Table of Contents