Maintenance
Carrying out the fresh air calibration
To improve the zero point accuracy, you can
carry out a fresh air calibration.
Calibrate the device to fresh air, free of
measured gases or other interfering
gases.
All the sensors are included in the fresh
air calibration. Sensors which have not
warmed up or which are faulty prevent a
calibration.
In the case of sensors which are in the
warm-up phase, the message » 159 « is
displayed with the special symbol » «
(for warning message).
In the case of a sensor or device error, the
message » 109 « is displayed with the
special symbol » « (for fault message).
The message is cleared after 5 seconds
and the function is available again in the
menu.
During the fresh air calibration, the zero
point of all the sensors (with the
exception of the DrägerSensor XXS O
2
)
is set to 0.
In the case of the DrägerSensor XXS O
2
,
the display is set to 20.9 vol. %.
Switch on the device.
Depending on device configuration:
Call the quick menu and select the "fresh
air calibration" function, page 15.
or
Call the calibration menu and select the
"fresh air calibration" function, page 16.
The current gas concentration values
flash.
Press the key to start the fresh air
calibration.
03123854
_
01
_
en.e
p
s
OK