Maintenance
Replacing the Batteries /
Rechargeable Batteries
Switching off the device:
Keep the key and the key
depressed at the same time.
1 Loosen the screw (2.5 mm hexagon
socket) on the power pack and remove
the power pack.
2 Replace the alkaline batteries with new
ones or rechargeable NiMHy batteries
with charged ones – ensure correct
polarity.
3 Completely replace the power pack T4
(with sealed rechargeable batteries,
order no. 83 18 704).
Insert the power pack into the device and
tighten the screw.
After replacing the power pack T4, it is
recommended that a complete charging is
carried out.
After batteries have been replaced:
The settings and data are stored when
the battery is replaced.
The sensors warm up again.
CAUTION:
Do not replace the batteries / rechargeable batteries in hazardous areas. Danger of
explosion!
Batteries / rechargeable batteries are part of the Ex approval. Only the following types
may be used:
Alkaline batteries – T4 – (not rechargeable!)
Energizer No. E91, Energizer No. EN91 (Industrial), Varta Type 4106 (power one)
NiMHy rechargeable batteries – T3 – (rechargeable)
GP 180AAHC (1800) max. 40
o
C ambient temperature.
Caution:
Do not throw used batteries into fire or
try to open them by force. Danger of
explosion!
Dispose of the batteries in accordance
with local regulations.