Operation
Operation
Switching on the device
Press and hold the key for approx. 3 seconds until the countdown »3.2.1« shown
in the display has elapsed.
All the display segments, including the visual, audible and vibration alarms, are activated
for a short time.
The software version is displayed.
The device performs a self test.
The remaining operating time is displayed, e.g. »d73 (remaining operating time
730 days). This applies to Dräger X-am 1700 and if the service life is activated in Dräger
X-am 2000.
All alarm setpoints A1 and A2 as well as » « (TWA)
1)
and » « (STEL)
*)
for H
2
S
and CO are displayed in succession.
The time until the calibration interval expires is displayed in days, e.g. » CAL 73 «.
The sensor which is next due for calibration/adjustment is displayed with the days remai-
ning until the next calibration/adjustment e.g. » Ex %UEG CAL 20 «.
During the warm-up time of the sensors, the respective display of the measured value flas-
hes and the special symbol » « (for warning) is displayed. No alarms are issued during
the running-in period of the sensors.
Press the key to cancel the display of the activation sequence.
Switching off the device
Press and hold the key and the key at the same time until the countdown
»3.2.1« shown in the display has elapsed.
Before the device is switched off, the visual, audible and vibration alarms are activated for
a short time.