Maintenance
Automatic implementation with the Bump Test Station
Prerequisite:
The device must first be configured for the automatic function test with gas (bump test) using
the PC software "CC Vision".
Activate the device for the automatic function test.
Composition of test gas (mixed gas) – standard on delivery: 50 ppm CO, 15 ppm H
2
S, 2.5
vol. % CH
4
, 18 vol. % O
2
Define which measuring channels should participate in the automatic function test. By de-
fault, all measuring channels participate in the function test.
Prepare the Bump Test Station according to the instructions.
Switch on the device and insert it into the receptacle of the Bump Test Station until it locks
into place.
The function test with gas is started auto-
matically. The special symbol » « (for
bump test) flashes.
If the gas concentration exceeds the
alarm setpoint A 1 or A 2, the relevant
alarm is triggered, which is automatically
acknowledged after approx. 10 seconds.
Next:
The display of the current gas concentra-
tion changes with the display »O.
The bump test that was carried out is do-
cumented with the result and date in the
device memory.