Maintenance
Calibrating/adjusting the sensitivity
for a single measuring channel
The span calibration can be carried out
specifically for individual sensors.
In the case of the span calibration, the
sensitivity of the selected sensor is set to
the value of the used test gas.
Use a standard test gas.
Allowed test gas concentration:
Connect the test gas cylinder with the ca-
libration cradle.
Vent the test gas leaving the adapter into
a fume cupboard or into the open air (with
a hose connected to the second connec-
tor of the calibration cradle).
Switch on the device and insert it into the
calibration cradle.
Press the [ + ] key and keep it depressed
for 5 seconds to call the calibration menu,
enter the password and select the "span
calibration" function, page 16.
Press the key to start the span cali-
bration.
The display flashes the gas of the first
measuring channel, e.g. » Ex - %LEL «.
Press the key to carry out the calibra-
tion of this measuring channel.
or
Use the key to select another measu-
ring channel (O
2
- vol. %, H
2
S - ppm or
CO - ppm).
Ex: 40 to 100 %LEL
O
2
10 to 25 vol. %
CO: 20 to 999 ppm
H
2
S: 5 to 99 ppm
CAUTION
Do not inhale the test gas. Risk to health!
Observe the hazard warnings of the rele-
vant Safety Data Sheets.
02723854_01_en.eps
0
0.5 L/min
03523854_01_en.eps
OK
OK