Maintenance
Press the key to carry out the calibra-
tion of the selected measuring channel.
The calibration gas concentration is dis-
played.
Press the [OK] key to confirm the cali-
bration gas concentration or use the [+]
key to change the calibration gas concen-
tration and complete the process by pres-
sing the [OK] key.
The measurement value flashes.
Open the test gas cylinder valve to let test
gas flow over the sensor.
The displayed, flashing measurement va-
lue changes to the value according to the
supplied test gas.
When the displayed value has stabilized
(after at least 120 seconds):
Press the key.
The display containing the current gas
concentration changes with the display
»O.
Press the key to quit the calibration of
this measuring channel.
The next measuring channel is offered for
calibration.
After the last measuring channel has
been calibrated/adjusted, the device swit-
ches to measuring mode.
Close the test gas cylinder valve and re-
move the device from the calibration cr-
adle.
If a fault has occurred during the span ca-
libration.
The fault message » « appears and
» « is displayed for the respecti-
ve sensor instead of the measured value
In this case, repeat the calibration.
If necessary, replace the sensor,
page 46.
Note on adjusting the Ex channel to
nonane as the measured gas:
When calibrating the Ex channel, pro-
pane can be used as a substitute for the
calibration gas.
When using propane to adjust the Ex
channel to nonane, the display must be
set to 2x the test gas concentration used.