27
Configuring the Device
Device settings
The following changes can be made to the device parameters for a device:
NOTICE
Only trained personnel are permitted to make changes to the device configuration.
Designation Field
Password Numeric field (3-figure)
Operating signal LED
1)
Yes/No
Operating signal horn
1)
Yes/No
Switch-off mode “Switch off permitted” or
“Switch off prohibited” or
“Switch off prohibited at A2”
Shift length (TWA)
2)
(in minutes) 60 - 14400
(setting for exposure alarm)
Short-term exposure limit (STEL)
3) 4)
(in
minutes)
0 - 15
(setting for exposure alarm)
User ID(12 characters) Alphanumeric field
Switch database on or off On/Off
Overwrite database Yes/No
Database mode Peak/Average
Database interval 1 s / 10 s / 30 s / 1 min / 2 min / 5 min /
10 min / 30 min
Date (date on the PC)
Time (time on the PC)
Warning after expiry of calibration interval Yes/No
Error after expiry of calibration interval Yes/No
Delay until error after expiry of calibration
interval (days)
0 - 10
Automatic detection of Bump Test Station Yes/No
Activate sensitivity calibration following
negative bump test
Yes/No (relates only to a device
connected to the Dräger Bump Test
Station)
Bump test mode “extended bump test” or “quick bump test”
or “bump test deactivated”
Warning after expiry of bump test interval Yes/No
Error after expiry of bump test interval (if
warning activated)
Yes/No
Capture range Yes / No
Remote configuration Yes / No
Bump test interval (days) 1 - 732