Faults, Cause and Remedy
Fault messages
656 The calibration interval for
DrägerSensor XXS EC 4 has
elapsed
Carry out span calibration for
DrägerSensor XXS EC 4,
page 50.
657 Alarm setpoint A2 setting is
greater than 60 %LEL
Set alarm setpoint to less than
60 %LEL.
Special symbol »«
and displayed numerical
code:
Cause Remedy
102 The customer's service life
counter has elapsed
Reset the service life counter
using Dräger CC Vision.
103 The device is defective The device must be repaired by
DrägerService.
104 Check sum error program
code
The device must be repaired by
DrägerService.
105 The bump test interval has
elapsed
Carry out bump test, page 43.
106 The calibration interval has
elapsed (at least 1 calibration
interval has elapsed)
Carry out span calibration,
page 47 or page 50.
107 Bump test error (at least
1 channel has a bump test
error)
Carry out bump test, page 43 or
carry out span calibration,
page 47 or page 50.
108 The device is defective The device must be repaired by
DrägerService.
109 The menu function cannot be
carried out because of an
error.
Determine the error code via the
info menu and switch it off, if
necessary.
111 Failed alarm element test:
alarm light.
Repeat alarm element test with
Dräger X-dock.
112 Failed alarm element test:
alarm horn.
Repeat alarm element test with
X-dock.
113 Failed alarm element test:
Vibration motor.
Repeat alarm element test with
X-dock.
114 Defective parameter check Correct parameters and repeat
test using X-dock
115 Device is disabled by X-dock. Activate device with X-dock.
116 Failed software update. The device must be repaired by
DrägerService.
Special symbol » «
and displayed numerical
code:
Cause Remedy