53
Maintenance
If a fault occurred during the span
calibration.
The fault message » « appears
and » « is displayed for the
sensor instead of the measured
value.
In this case, repeat the calibration.
If necessary, replace the sensor,
page 59.
If the gas CH
4
(measurement range
up to 100 vol. %) is selected on the
CatEx channel, the heat conductivity
is calibrated, display with
channelselection:
Press the key to perform the
calibration for heat conductivity.
Press the key to select the next
sensor.
Press the key to perform the
calibration of the selected measuring
channel.
The test gas concentration is
displayed.
Press the key to confirm the test
gas concentration or use the [+] key
to change the test gas concentration
and complete the process by
pressing the key.
The measurement value flashes.
Open the test gas cylinder valve to let
test gas flow over the sensor.
The displayed, flashing measurement value changes to the value according to the
supplied test gas.
Press the key to select the next sensor.
03723999_01_en.eps
ch
4
%LEL
CO
2
ppm
O
2
Vol%
CO
ppm
H
2
S
ppm
NH
3
ppm
05123999_01_en.eps
CH
4
VOL%
CO
2
ppm
O
2
Vol%
CO
ppm
H
2
S
ppm
NH
3
ppm
OK
05223999_01_de.eps
O2
VOL%
O
2
Vol%
CO
ppm
H
2
S
ppm
NH
3
ppm
OK
OK
OK