64
Technical Data
Technical Data
X-am 5000
Ambient conditions:
during operation and
storage
Temperature class T4 (-20 to +50 °C):
NiMH power packs type: HBT 0000, HBT 0100
Power pack type: ABT 0100
with alkaline single cell type: Duracell Procell MN 1500
1)
,
Duracell Plus Power MN 1500
1)
Temperature class T3 (-20 to +40 °C):
Power pack type: ABT 0100
with NiMH single cell type: GP 180AAHC
1)
with alkaline single cell type: Panasonic LR6 Powerline
Temperature class T3 (0 to +40 °C):
Power pack type: ABT 0100
with alkaline single cell type: Varta 4006
1)
, Varta 4106
1)
1) Not part of the BVS 10 ATEX E 080 X and PFG 10 G 001 X technical suitability tests.
Temperature range over
a short period (ATEX &
IECEx only)
1)
:
-40 to +50 °C
Maximum of 15 minutes with NiMH power pack T4 (HBT
0000) or T4 HC (HBT 0100)
Prerequisite: storage of the instrument at room temperature
(+20 °C) for at least 60 minutes in advance.
Air pressure 700 to 1300 hPa
Humidity 10 to 90 % (to 95 % briefly) relative humidity
Device data
Protection class IP 67 for devices with sensors
Alarm volume Typically 90 dB (A) in 30 cm distance
Operation time
Alkaline battery Typically 12 hours under normal conditions
– NiMH
power pack:
T4 (HBT 0000)
T4 HC (HBT 0100)
Typically 12 hours under normal conditions
Typically 13 hours under normal conditions
Dimensions approx. 130 mm x 48 mm x 44 mm (H x W x D)
Weight approx. 220 g to 250 g
Storage time
X-am 5000
Sensors
1 year