Fault messages
Special symbol
»« and displayed
numerical code:
Cause Remedy
102 The customers service
lifecounter has elapsed
Reset the service life counter
using Dräger CC Vision.
103 The instrument is defective The instrument must be
repaired by Service.
104 Check sum error program
code
The instrument must be
repaired by Service.
105 The bump test interval has
elapsed
Carry out bump test, page 45.
106 The calibration interval has
elapsed (at least 1 calibration
interval has elapsed)
Carry out span calibration/
adjustment, page 50 or
page 53.
107 Bump test error (at least 1
channel has a bump test
error)
Carry out bump test, page 45 or
carry out span calibration/
adjustment, page 50 or page 53.
108 The instrument is defective The instrument must be
repaired by Service.
109 The menu function cannot be
carried out because of an
error.
Determine the error code via the
info menu and switch it off, if
necessary.
111 Failed alarm element test:
alarm light.
Repeat alarm element test with
Dräger X-dock.
112 Failed alarm element test:
alarm horn.
Repeat alarm element test with
X-dock.
113 Failed alarm element test:
Vibration motor.
Repeat alarm element test with
X-dock.
114 Defective parameter check Correct parameters and repeat
test using X-dock
115 Device is disabled by X-dock. Activate device with X-dock.
116 Failed software update. The device must be repaired by
DrägerService.
117 User parameters not feasible Check configuration of user
parameters and adjust
118 Flow alarm of X-am Pump Check the gas circuit for
obstructions and replace filters if
necessary.