EasyManua.ls Logo

Dräger X-am 5600 - Calibrating;Adjusting the Instrument

Dräger X-am 5600
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
47
Calibrating/adjusting the instrument
Calibration may not be possible due to instrument and channel errors.
Allow the sensors to warm up before the calibration!
Warm-up time: refer to the Instructions for Use/data sheets of the Dräger Sensors
installed (on CD).
Calibration interval:
Observe the relevant specifications in the Instructions for Use/data sheets of the
Dräger Sensors installed.
For critical applications, observe the recommendations in EN 60079-29-2
1)
or
EN 45544-4
2)
and national regulations. We recommend that you calibrate the
channels after 6 months.
Improve the zero point accuracy – carry out the fresh air calibration/zero-point
calibration, page 48.
Set the sensitivity of all sensors to the value of the test gas – carry out the 1-button
calibration, page 50.
Set the sensitivity of a sensor to the value of the test gas – span calibration/
adjustment, page 53.
NOTICE
A potentially activated H
2
added signal is automatically temporarily deactivated
during a manual calibration, a PC calibration or an automatic Bump Test for the
relevant duration.
NOTICE
Dräger recommends using the extended bump test for cross calibrations
(Dräger X-dock technical manual).
1) EN 60079-29-2 – Guidelines for selection, installation, use and maintenance of instruments for
the detection and measurement of flammable gases and oxygen.
2) EN 45544-4 – Electrical instruments for the direct detection and direct concentration
measurement of toxic gases and vapours – Part 4: Guidelines for selection, installation, use and
maintenance.
WARNING
Risk to health! Do not inhale the test gas.
Observe the hazard warnings of the relevant Safety Data Sheets.

Table of Contents

Other manuals for Dräger X-am 5600

Related product manuals