47
Calibrating/adjusting the instrument
Calibration may not be possible due to instrument and channel errors.
Allow the sensors to warm up before the calibration!
Warm-up time: refer to the Instructions for Use/data sheets of the Dräger Sensors
installed (on CD).
Calibration interval:
— Observe the relevant specifications in the Instructions for Use/data sheets of the
Dräger Sensors installed.
— For critical applications, observe the recommendations in EN 60079-29-2
1)
or
EN 45544-4
2)
and national regulations. We recommend that you calibrate the
channels after 6 months.
— Improve the zero point accuracy – carry out the fresh air calibration/zero-point
calibration, page 48.
— Set the sensitivity of all sensors to the value of the test gas – carry out the 1-button
calibration, page 50.
— Set the sensitivity of a sensor to the value of the test gas – span calibration/
adjustment, page 53.
NOTICE
A potentially activated H
2
added signal is automatically temporarily deactivated
during a manual calibration, a PC calibration or an automatic Bump Test for the
relevant duration.
NOTICE
Dräger recommends using the extended bump test for cross calibrations
(Dräger X-dock technical manual).
1) EN 60079-29-2 – Guidelines for selection, installation, use and maintenance of instruments for
the detection and measurement of flammable gases and oxygen.
2) EN 45544-4 – Electrical instruments for the direct detection and direct concentration
measurement of toxic gases and vapours – Part 4: Guidelines for selection, installation, use and
maintenance.
WARNING
Risk to health! Do not inhale the test gas.
Observe the hazard warnings of the relevant Safety Data Sheets.