Calibrating/adjusting the sensitivity for an individual measuring channel
— The span calibration/adjustment can be carried out specifically for individual
sensors.
— In the case of the span calibration/adjustment, the sensitivity of the selected sensor
is set to the value of the test gas used.
— Use a standard test gas.
Allowed test gas concentration:
Connect the test gas cylinder with the
calibration cradle.
Vent the test gas into a fume
cupboard or into the open air (with a
hose connected to the second
connector of the calibration cradle).
_________________________________
1) Depending on the selected data set.
2) Depending on the measuring range and the measuring accuracy.
DUAL IR Ex
IR Ex
20 to 100 %LEL
1)2)
/ 5 to 100
1)2)
vol. %
DUAL IR CO
2
IR CO
2
0.05 to 5 vol.-%
2)
O
2
10 to 25 vol.-%
CO: 20 to 999 ppm
H
2
S: 5 to 99 ppm
H
2
HC 0,5 to 4,0 vol.-%
NO
2
5 to 99 ppm
Test gas concentration of other gases:
see Instructions for Use for the respective Dräger sensors.
WARNING
Risk to health!
Do not inhale the test gas.
Observe the hazard warnings of the
relevant Safety Data Sheets.