62
Replacing the Sensors
To replace the sensors of the instrument, connect the instrument with a PC.
Replace the sensors using the PC program Dräger CC Vision.
With each replacement of the DrägerSensors DUAL IR Ex/CO
2
, IR Ex or IR CO
2
,
the foam ring located above the sensor should also be exchanged.
CAUTION
Damage to components!
There are components in the instrument that are sensitive to electric charge. Before
opening the instrument to replace the sensor, ensure that the person performing the
work is earthed to avoid damage to the device. Earthing can be safely ensured, e. g.
via an ESD workstation (electrostatic discharge).
NOTICE
To open the instrument, unfasten the casing screws using a screwdriver (Torx T6).
IR
Calibration cradle (order no. 83 18 752)
with inserted
USB DIRA with USB cable (order no. 83 17 409)
USB DIRA with USB cable
(order no. 83 17 409)
E-Cal module Dräger X-am 125
(order no. 83 18 754)
USB 2.0
USB 2.0
USB 2.0 / COM
01433072_01_en.eps
Dräger
CC-Vision
Dräger
CC-Vision
Dräger
CC-Vision
E-Cal
0
0
0
X-am 5600