Maintenance
140 X-plore 2100
4.2.1 Cleaning and disinfecting procedure
1. Remove any accessories from the facepiece and clean them
separately
1)
.
2. Prepare a cleaning solution consisting of water and a cleaning
agent.
3. Clean all parts using a soft cloth with cleaning solution.
4. Rinse all parts thoroughly under running water.
5. Prepare a disinfecting solution consisting of water and a
disinfecting agent.
6. Place all parts that need to be disinfected into the disinfectant bath.
7. Rinse all parts thoroughly under running water.
8. Allow all parts to dry in the open air or in a drying cabinet (max.
temperature: 140
°F/60 °C). Do not expose to direct sunlight. Do
not shake the facepiece in order to remove water.
9. Mount the accessories if necessary.
4.2.2 Alternative cleaning and disinfecting procedure
1. Remove any accessories from the facepiece and clean them
separately
2)
.
2. Prepare a solution consisting of water and a combined cleaning
and disinfecting agent.
3. Immerse the facepiece in the solution. Make sure that all parts are
wetted.
4. Rinse all parts thoroughly under running water.
5. Allow all parts to dry in the open air or in a drying cabinet (max.
temperature: 140
°F/60 °C). Do not expose to direct sunlight. Do
not shake the facepiece in order to remove water.
6. Mount the accessories if necessary.
1) Strictly follow the corresponding instructions for use.
2) Strictly follow the corresponding instructions for use.