2) POWER B’D Removal/Installation
* CAUTION : Use a grounded strap when handling any electronic components.
- The user can’t fix the POWER BOARD. There will be malfunction or damage if the user fixes it
arbitrarily.
- Please replace the power board in case of BOARD fails.
- Please refer to 3. Assembly view when main board disassembles.
- The installation is in reverse order of disassemble view.
10. Attaching the tubing and user guide
- Attach sleeve and calibrated pressure gauge to the controller.
- Set the pressure value to 40mmHg on the controller.
- Check the pressure value visually by turning on the controller.
- Refer below check point of the graph for pressure value.
- The error value of setting pressure is ±15%.
- The standard chamber of setting pressure is 2nd chamber (Middle chamber)
* LEG Sleeve
Setting pressure
1ch
2ch
3ch
40mmHg
45mmHg
(51 ~ 39mmHg)
40mmHg±15%
(46 ~ 34mmHg)
35mmHg
(41 ~ 29mmHg)
* FOOT Sleeve
Setting pressure
1ch
120mmHg
120mmHg±15%
(138 ~ 102mmHg)
Sleeve
Main
controller
Sleeve
Pressure
gauge