11. Controller Maintenance and test methods
11-1. Test mode Set-up
When you press 'LEG' and 'MODE' buttons for 2 seconds at the same time, test method mode
including 10 test programs starts. Whenever you press the 'mode' button, the test number is
changed from 1 to 10 and when you press the 'START', you can change the selected value or do
the test. If you want to finish test method, please press 'START' for 3 seconds.
11-2. Test mode table
Check the program version
Check voltage for charging battery
Check pressure sensor or calibration
- The version indication of controller’s
software
- Left : MAIN Version
- Right : POWER Version
LYMPH MODE can be deleted if DVT
MODE is only used
- When you press “START”. You can
turn on out the ‘LYMPH’ mode.
- Check the supply voltage of controller
- If the supply voltage of controller is
below the minimum voltage, the
recharge may not be completed.