EasyManua.ls Logo

Eberline ESP-1 - Page 24

Default Icon
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The
calibration
constant
will
have
to
be
changed
when
switching
detectors.
When
changing
back
from
another
detector
to
the
detector
which
was
purchased
with
the
ESP-1,
use
the
calibration
constant
which
is
given
on
the
calibration
sheet
supplied
with
the
combined
ESP-1
and
detector.
Change
values
by
pressing
simultaneously
either
RESET
and
+
or
RESET
and
-.
Selection
of
Calibration
Constant
for
various
detectors:
(1)
If
the
detector
was
not
purchased
with
the
ESP-1
and,
thus,
a
calibration
sheet
is
not
available,
use
the
nominal
value
for
the
particular
Eberline
detector
which
is
given
in
section
V,
"Maintenance,"
11292-A36.
(2)
To
increase
the
value
of
CC,
press
RESET
and
+
simultaneously.
To
decrease
the
value
of
CC,
press
RESET
and
-
simultaneously.
(3)
To
calculate
a
calibration
constant
for
detectors
other
than
those
listed
in
the
table,
the
sensitivity
of
the
detector
must
be
known
and
is
usually
found
in
the
list
of
specifications
given
on
the
catalog
sheet.
Calculations
using
a
HP-270
detector
as
an
example
are
given
as
follows:
sensitivity
=
1200
c/min/mR/h
Calibration
=
1200
c/min
x
1000
mR/R
x
60
min/h
=
7.20
+
07
c/R
mR/h
4.
Setting
the
Dead
Time
(DT)
a.
Definition:
The
dead-time
correction
constant
is
a
derived
number
used
to
correct
for
counting
losses
due
to
inability
of
the
detector
to
recover
at
high
counting
rates.
This
correction
results
in
a
more
linear
response
to
the
radiation
field
being
measured
and
extends
the
useful
range
of
some
detectors
used
with
the
ESP-1
by
a
factor
of
as
much
as
ten
times
(which
provides
the
equivalent
of
an
extra
range
on
a
standard
rate
meter).
Selection
of
"DT"
for
detectors
which
were
purchased
from
the
factory
with
the
instrument:
If
the
ESP-1
was
purchased
with
a
detector
from
the
factory,
the
DT
value
will
be
correctly
set
at
the
factory
and
the
next
section
may
be
bypassed
until
it
is
necessary
to
use
a
different
detector
or
until
time
for
routine
calibration.
When
changing
back
from
another
detector
to
the
detector
purchased
with
the
ESP-1,
use
the
DT
given
on
the
calibration
shect
supplied
with
the
combined
ESP-1
and
detector.
ESPI.MAN/February
1992

Related product manuals