EasyManua.ls Logo

Eberline RO-2 - Instrument Operation; Description of Controls; Using the Instrument and Important Notes

Default Icon
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
`
*
-
MODEL
RO-2
SECTION
II
OPERATION
A.
DESCRIPTION
OF
CONTROLS
1.
Function
Switch:
Eight-position
rotary
switch
that
turns
the
instrument
OFF,
checks
the
condition
of
the
batteries,
checks
instrument
ZERO,
and
selects
the
range
of
operation
to
be
used.
2.
ZERO
Knob:
Used
to
set
the
meter
to
zero
when
the
ZERO
switch
position
is
selected
or
when
in
an
insignificant
radiation
field.
3.
Calibration
Controls:
Four
variable
resistors,
one
for
each
range.
.
B.
USING
THE
INSTRUMENT
1.
Turn
the
function
switch
to
ВАТ
/
and
then
to
BAT
2
positions.
The
meter
should
read
above
the
BATT
cut-off
line
in
both
cases.
2,
Turn
the
function
switch
to
ZERO
position.
Check
that
the
meter
reads
zero.
If
not,
set
it
to
zero
with
the
ZERO
knob.
a.
Due
to
a
change
in
manufacturing
of
the
oper-
ational
amplifier,
Al,
the
zero
control
will
work
in
reverse
for
instruments
with
the
new
manufacturer's
amplifier.
These
devices
can
.
be
identified
by
the
description
“DAWN”
indicated
on
the
part.
This
should
pose
no
problem,
simply
turn
the
zero
control
in
either
direction.
To
zero
the
meter
as
appropriate.
3.
Set
the
function
switch
to
the
desired
range
of
operation.
The
switch
position
selected
is
the
full
scale
reading
of
that
range.
4.
When
measuring
8,
low
energy
у,
ог
x-ray
emis-
sions,
open
the
sliding
8
shield
on
the
bottom
of
the
case
_and
face
the
bottom
of
the
instrument
toward
the
radia-
tion
source.
To
open
or
close
the
shield,
depress
the
Їгїс-
CHANGE
3
tion
release
button
on
the
left
side
of
the
case
and
manually
move
the
slide,
or
let
it
fall
due
to
gravity.
When
the
shield
is
open,
protect
the
thin face
against
puncture
damage.
NOTES
a.
The
zero
setting
of
the
instrument
may
be
checked
in
any
radiation
field
by
merely
selecting
the
ZERO
position.
b.
When
selecting
the
most
sensitive
range
(5
mR/h),
switching
transient
noise
may
cause
a
tem-
porary
deflection
of
the
meter.
This
can
be
avoided
by
first
selecting
50
mR/h,
letting
the
needle
settle,
and
then
switching
to
5
mR/h.
с.
The
effective
center
of
the
ion
chamber
is
marked
by
dimples
at
the
front
and
sides
of
the
instru-
ment
case.
d.
Since
the ion
chamber
is
vented
to
atmospheric
pressure,
it
is
sensitive
to
changes
in
both
air
pressure.
and
temperature.
Tables
2-1
and
2-2
give
correction
fac-
tors
to
be
used
if
the
use
conditions
are
different
from
the
calibration.
conditions.
If
both
pressure
and
tem-
-
perature
are
different,
multiply
the
meter
reading
by
both
factors.
е.
Do
not
expose
the
instrument
to
fields
significantly
above
full
scale
for
the
range
selected
or
amplifier
gate
stress
and
possible
damage
may
result.

Related product manuals