HYDRA-DS NO
3
-N Page 47
S80 Serial Number (hi word)
Unit Serial Number (32 bit integer hi word)
S80 Serial Number (lo word)
Unit Serial Number (32 bit integer lo word)
Specific S80 sensor type (see S80 Sensor Types tab)
Specific chemicals the S80 is set to detect (see S80
Sensor Types tab)
Max sensor range (bytes 3 and 2)
Max sensor range (bytes 1 and 0)
Min sensor range (bytes 3 and 2)
Min sensor range (bytes 1 and 0)
S80 Sensor Value (hi word)
Current sensor value (bytes 3 and 2)
S80 Sensor Value (lo word)
Current sensor value (bytes 1 and 0)
S80 Sensor Voltage (hi
word)
Corresponding sensor voltage to the sensor value
(byte 3 and byte 2)
S80 Sensor Voltage (lo
word)
Corresponding sensor voltage to the sensor value
(byte 1 and byte 0)
S80 Sensor Temperature (hi
word)
Sensor Temperature (bytes 3 and 2)
S80 Sensor Temperature (lo
word)
Sensor Temperature (bytes 1 and 0)
Sensor is a Sentinel Type (0 - No, 1 - Yes)
% of Sensor life remaining
S80 Sentinel Vs (hi word)
Scaled Sentinel Voltage (in mV) normalized to Vo
(bytes 3 and 2)
S80 Sentinel Vs (lo word)
Scaled Sentinel Voltage (in mV) normalized to Vo
(bytes 1 and 0)
S80 Sentinel Vo (hi word)
Sentinel 100% value (in mV) on the life relative to 0V
(bytes 3 and 2)
S80 Sentinel Vo (lo word)
Sentinel 100% value (in mV) on the life relative to 0V
(bytes 1 and 0)
S80 Sentinel Range (hi word)
Sentinel Range (bytes 3 and 2)
S80 Sentinel Range (lo word)
Sentinel Range (bytes 1 and 0)
Sensor Full Name (18
characters max)
ASCII character bytes 0 and 1, ex. "AB" A - 65 (41
hexidecimal), B - 66 (42 hex), send 6566 (4142 hex).
The characters permitted are space ' ' (32 base 10, 20
hex) through '}' 125 base 10, 7D hex).
Sensor Abbreviated Name (8
characters max)
ASCII character bytes 0 and 1, ex. "AB" A - 65 (41
hexidecimal), B - 66 (42 hex), send 6566 (4142 hex).
The characters permitted are space ' ' (32 base 10, 20
hex) through '}' 125 base 10, 7D hex).