12.3
Error codes of the DryExxEco control
®
Advanced-ETH
The error codes are displayed in plain text on the screen of the control
DryExxEco
®
Advanced-ETH.
Fault description Cause Remedy
Screen output:
SI automated circuit
breaker, dig. outputs
triggered!
Excess current
Localise and rectify the cause of the excess
current. Switch the fuse in the switch cabinet
back on.
Short circuit Localise and rectify the cause of the short
circuit. Switch the fuse in the switch cabinet
back on.
Screen output:
Error: SI automated circuit
breaker
, cooler has
triggered!
Excess current
Localise and rectify the cause of the excess
current. Switch the fuse in the switch cabinet
back on.
Short circuit Localise and rectify the cause of the short
circuit. Switch the fuse in the switch cabinet
back on.
Screen output:
Fault: Belt lubricant empty
Belt lubricant
supply empty
Provide new belt lubricant (
Ä
Chapter
8.3.2 „Failure of the belt lubricant - empty
message“ on page 47).
Screen output:
Fault: Communication with
partner (enable)
Ethernet
communication is
interrupted.
Localise and rectify communication break.
The following fault can occur in all circuits (1-16), but is only described here
once as an example.
This error message can only be displayed, if in
Ä
Chapter 10.8.2 „Overview:
[General parameters] - [operating mode]“ on page 74 the monitoring of the
dosing quantity is switched on (button [ON] green)!
Fault description Cause Remedy
Screen output:
Fault: Dosing quantity circuit 1 =>
check plant => ECOLAB
Overdosing, e.g. due to line
break or missing spray
nozzle.
Localise and replace
line break or missing
spray nozzle.
Underdosing blocked line
system or nozzle.
Clean, or replace,
blocked line system or
nozzle.
Pump is conveying with the
incorrect values
Check the pump.
Troubleshooting
Error codes of the DryExxEco control® Advanced-ETH
86Rev. 02-09.2017