417101774_EMP_KKS_E60_E.doc - 50 - Rev. 8-03.10
10.3.14.7 MicroFlow
PLUS
/ Calibration
Use the "Calibration" option to newly calibrate the transducer during initial installation or
when changing the product.
The pump must be running and the metering medium must flow through the transducers
without any bubbles before starting the calibration process.
On activating the “calibration” function, the values for min. conductivity (electrical connection
between the sensor pegs in the measuring sensor disrupted) und max. conductivity
(electrical connection) are determined.
These values are then used to calculate and display a difference.
To monitor the system well, the difference should be greater than 3..4.
Fig. 10.45
10.3.15 Configuration / Batch
NOTE
The pump must be calibrated before using batch metering (see chapter 10.4).
When activating batch mode with a signal to the metering lock/pulse input (plug II), a
previously defined quantity is metered with 100 % stroke frequency.
The batch metering can be interrupted by deactivating the release (metering lock) or
switching off the pump.
10.3.15.1 Selecting
- Batch is inactive
(Default setting)
9 Batch is active
Fig. 10.46
If batch metering is active, the
following items appear in the
submenu:
Menu item Chapter
Quantity 10.3.15.2
Input 10.3.15.3
10.3.15.2 Batch / Quantity
“Quantity” is used to est
ablish the desired metering quantity (in µl, ml, l or G) per batch
metering process. Setting range:
0 – 99999